引用本文: | 王沫楠,王红晶,杜志江.股骨假体热应力有限元分析[J].哈尔滨工业大学学报,2012,44(7):38.DOI:10.11918/j.issn.0367-6234.2012.07.007 |
| WANG Mo-nan,WANG Hong-jing,DU Zhi-jiang.Finite element analysis of thermal stress of femoral prosthesis[J].Journal of Harbin Institute of Technology,2012,44(7):38.DOI:10.11918/j.issn.0367-6234.2012.07.007 |
|
摘要: |
为定量研究材料热物性对股骨-假体系统应力的影响,以热弹性力学为基础,利用有限元方法对髋关节置换后股骨-假体系统的应力进行分析.假设假体、骨水泥、股骨为线弹性和各向同性材料且完全结合的同心圆柱体,经力学分析和有限元仿真,分别求得钛合金和钴铬合金作为假体材料时考虑热效应前后的假体/骨水泥、骨水泥/股骨随轴向位置分布的界面剪切应力以及假体、骨水泥、股骨的轴向应力.结果表明:界面失效是股骨-假体系统的主要失效形式,热效应使界面产生了混合而不是纯剪切的失效形式,加快了股骨-假体系统的界面失效,加速了假体柄的松动;由于钴铬合金的热膨胀系数大于钛合金,钴铬合金作为假体材料时更易引起界面失效,导致手术失败.选择热膨胀系数小的假体材料,可以减小界面失效的可能性,提高手术成功率. |
关键词: 股骨 假体 钛合金假体;钴铬合金假体;热应力 有限元分析 |
DOI:10.11918/j.issn.0367-6234.2012.07.007 |
分类号:R496 |
基金项目:国家自然科学基金资助项目(61073144); 黑龙江省新世纪优秀人才培养计划资助项目(251NCET006); 机器人技术与系统国家重点实验室开放研究资助项目. |
|
Finite element analysis of thermal stress of femoral prosthesis |
WANG Mo-nan,WANG Hong-jing,DU Zhi-jiang
|
1.State Key Laboratory of Robotics and System, Harbin Institute of Technology, 150080 Harbin, China;2.College of Mechanical and Power Engineering, Harbin University of Science and Technology, 150080 Harbin, China
|
Abstract: |
To study the influence of thermo property on stress of femur-prosthesis system quantitatively, the finite element method is used to analyze the stress of femur-prosthesis system when the hip joint has been replaced, in which the prosthesis, bone cement and femur are assumed as linear elastic and isotropic materials in shape of a concentric cylinder. Considering the thermal effect, the axial stress of prosthesis, bone cement, femur and the interface shear stress along axial of the prosthesis/bone cement and bone cement/femur when the prosthesis are made of titanium alloy and cobalt-chromium alloy are acquired separately by mechanical analysis and FEM simulation.The results indicate that the interface failure is the primary failure mode of the femur-prosthesis system. The existence of thermal effect accelerates the interface failure of femur-prosthesis system and the loose of prosthesis stem. The possibility of interface failure which result in operation failure will ascend when the cobalt-chromium alloy is used, because the thermal expansion coefficient of cobalt-chromium alloys is higher than that of titanium alloy. |
Key words: femur prosthesis titanium prosthesis cobalt-chromium prosthesis thermal stress finite element analysis |